| List of calibration services |
no. |
Test equipment to be calibrated or measured value/ quantity |
Measured or reproduced value or range using our standards |
Best measuremnet capability |
Expanded uncertainity |
Remarks |
1. 1. |
Reproduction of insulation resistance (VBK-1 procedure) |
10 kohm ... 30 Gohm |
0,6 % |
k = 2 |
BMC referred to measures value |
1. 2. |
Reproduction of insulation resistance (VBK-2 procedure) |
10 kohm ... 1 Tohm |
0,12 % |
k = 2 |
BMC up to 5 Mohm |
2. |
Measuring of insulation resistance test voltage |
50 V ... 1000 V |
64 ppm |
k = 2 |
BMC at 1000 V |
3. 1. |
Reproduction of low resistances (VBK-1) |
0 ohm ... 1,9 kohm |
0,6 % |
k = 2 |
BMC referred to measures value |
3. 2. |
Reproduction of low resistances (VBK-2) |
0 ohm ... 1 kohm |
0,17 % |
k = 2 |
BMC at 1000 ohm |
4. 1. |
Reproduction of line and loop resistance (VBK-1) |
0 ohm ... 1 kohm |
0,6 % |
k = 2 |
BMC referred to measures value |
4. 2. |
Reproduction of line and loop resistance and PE continuity (VBK-2) |
0 ohm ... 1 kohm |
0,6 % |
k = 2 |
BMC at 1000 ohm |
5. 1. |
Reproduction of earth resistance (VBK-1) |
||||
a) four lead method |
0 ohm ... 1,9 kohm |
0,6 % |
k = 2 |
BMC referred to measures value |
|
b) using one clamp |
0 ohm ... 1,9 kohm |
0,6 % |
|||
c) using two clamps |
0 ohm ... 290 ohm |
0,6 % |
|||
5. 2. |
Reproduction of earth resistance (VBK-2) |
||||
| a) four lead method | 0 ohm ... 1 kohm |
0,6 % |
k = 2 |
BMC at 1000 ohm |
|
| b) using one clamp | 0 ohm ... 1 kohm |
0,6 % |
BMC at 1000 ohm |
||
| c) using two clamps | 0 ohm ... 100 ohm |
0,6 % |
BMC at 100 ohm |
||
6. 1. |
RCD measurements (VBK-1) | ||||
| a) contact voltage | 0 V ... 125 V |
0,6 % |
k = 2 |
BMC referred to measures value |
|
b) trip out time |
18,1 ms ... 1711 ms |
2,2 ... 19 ms |
Time domain according t EN 61009 |
||
6. 2. |
RCD measurements (VBK-2) |
||||
| a) contact voltage | 0 V ... 125 V |
0,6 % |
k = 2 |
BMC referred to measures value |
|
| b) trip out time | 18,1 ms ... 5 s |
0,7 ms |
BMC referred to measures value |
||
| c) tripping current | 3 mA ... 3 A |
1,5 % |
BMC at 100 mA |
||
7. 1. |
Reproduction of varistor over-voltage protection (VBK-1) | 0 V ... 1000 V |
0,6 % |
k = 2 |
BMC referred to measures value |
7. 2. |
Reproduction of varistor over-voltage protection (VBK-2) | 0 V ... 1000 V |
0,12 % |
k = 2 |
BMC referred to measures value |
8. |
Voltage drop |
0 V ... 10 V |
0,2 % |
k = 2 |
BMC referred to measures value |
9. |
Leakage current | 2 mA ... 7,7 mA |
1,7 % |
k = 2 |
BMC at 7,7 mA |
10. |
Measurement of withstanding voltage |
||||
| a) voltage | 100 V ... 12 kV |
0,9 % |
k = 2 |
BMC at 12 kV |
|
| b) current | 50 µA ... 20 mA |
5 % |
BMC at 20 mA |
||
11. |
Reproduction of DC voltage |
0 mV ... 1020 V |
32 ppm |
k = 2 |
BMC at 20 V |
12. |
Reproduction of DC current |
0 µA ... 20,2 A |
81 ppm |
k = 2 |
BMC at 20 mA |
13. |
Reproduction of DC current for clamps |
0 A ... 1100 A |
0,58 % |
k = 2 |
BMC at 1000 A |
14. |
Reproduction of AC voltage |
0 mV ... 1020 V |
0,044 % |
k = 2 |
BMC at 20 V |
15. |
Reproduction of AC current |
0 µA ... 20,2 A |
0,12 % |
k = 2 |
BMC at 2 A |
16. |
Reproduction of AC current for clamps |
0 A ... 1100 A |
0,3 % |
k = 2 |
BMC at1000 A |
17. |
Reproduction of power |
0 W ... 1000 kW |
0,16 % |
k = 2 |
BMC at 400 W |
18. |
Reproduction of four wire resistance |
0 ohm ... 1 Gohm |
0,046 % |
k = 2 |
BMC at 1 kohm |
19. |
Reproduction of frequency |
10 Hz ... 10 MHz |
14 ppm |
k = 2 |
BMC referred to measures value |
20. |
Reproduction of capacity |
1 nF ... 100 µF |
0,31 % |
k = 2 |
BMC at 100 nF |
21. |
Measurement of DC voltage |
0 mV ... 1000 V |
0,004 % |
k = 2 |
BMC at 10 V |
22. |
Measurement of DC current |
0 mA ... 3 A |
0,07 % |
k = 2 |
BMC at 100 mA |
23. |
Measurement of AC voltage |
0 V ... 750 V |
0,1 % |
k = 2 |
BMC at 100 V |
24. |
Measurement of AC current |
0 mA ... 3 A |
0,5 % |
k = 2 |
BMC at 100 mA |
25. |
Measurement of four wire resistance |
0 ohm ... 100 Mohm |
0,01 % |
k = 2 |
BMC at 100 kohm |